high-frequency linear ultrasound probe

high-frequency linear ultrasound probe

¥12,000.00
跳至产品信息
high-frequency linear ultrasound probe

high-frequency linear ultrasound probe

I. Basic Information


Brand & Model: GE Healthcare L6-12-RS High-Frequency Linear Array Ultrasound Transducer

REF Number: 5459829 (Corresponding OEM Part Number: 5454332)

Manufacturing Date: April 2021

Place of Origin: China (Assembled by GE Healthcare in Wuxi)

Protection Rating: IPX7 (Immersion disinfection available)

II. Core Technical Specifications

 

Transducer Type:Linear Array
Frequency Range:4.0–12.0/13.0 MHz (Broadband coverage)
Element Count;128 elements
Scan Field of View:38.4–47 mm aperture
Maximum Scan Depth:Approximately 6 cm
Compatible Systems:GE ultrasound systems including LOGIQ Series (V2, P7, F, V Series), Vivid Series (T8/T9), Versana Series (Active/Premier) and other models

III. Core Advantages & Clinical Applications


High-resolution Superficial Imaging


With high frequency and optimized linear array structure, it presents clear details of superficial structures such as thyroid, breast, lymph nodes, tendons and ligaments. It supports Color Doppler and Pulsed-Wave Doppler functions.

Strong Compatibility & Wide Application Scenarios


It is applicable to examinations of small organs, musculoskeletal ultrasound, peripheral vascular assessment, pediatric superficial tissue imaging and interventional puncture guidance.

Ergonomic Design


The lightweight handle and narrow array design enable easy access to anatomical blind spots like supraclavicular fossa and posterior knee area, enhancing comfort and convenience during examinations.

High Durability & Compliance


Equipped with medical-grade housing and anti-corrosion gold-plated connectors. The IPX7 waterproof rating fits conventional hospital disinfection protocols and lowers the risk of damage.

IV. Usage and Maintenance Notes


  • Compatibility: Only compatible with GE ultrasound systems with matching interfaces. Please confirm the system model and firmware version before use.
  • Disinfection Requirements: IPX7 rating allows immersion disinfection. Follow the user manual for specified disinfectant type, concentration and immersion time.
  • Daily Maintenance: Avoid excessive bending of the transducer cable. Clean off coupling gel timely after use to prevent hardening which may affect imaging performance.